QuickDetect™ CD163 (Human) ELISA Kit

A rapid sandwich ELISA kit for the quantitative measurement of CD163 in human serum, plasma, cell culture supernatnat and other biological fluids

WARNING: This product can expose you to chemicals including TMB, which is [are] known to the State of California to cause cancer.  For more information go to www.P65Warnings.ca.gov.
Catalog #: K4422 | abID:

Product Details

Cat # +Size K4422-100
Size 100 assays
Detection Method Absorbance (450 nm)
Species Reactivity Human
Applications This ELISA kit is used for quantitative measurement of CD163 in human serum, plasma, cell culture supernatant and other biological fluids.
Features & Benefits • Easy, convenient and time-saving method to measure the level of CD163 in human serum, plasma, culture supernatants and other biological fluids.
The entire assay can be performed within 90 minutes with only a single incubation and a wash step respectively, resulting in fewer handling steps, fewer errors and more consistent results.
Detection Range: 9.3 - 300 ng/ml

• Cross Reactivity: No significant cross-reactivity or interference between this analyte and its analogues was observed.
Kit Components • Micro ELISA Plate
• Standard
• Standard Diluent
• Special Diluent
• HRP-Conjugate Reagent
• Chromogen Solution A
• Chromogen Solution B
• Stop Solution
• Wash buffer (20X)
• Sealed bags
• Microplate Sealers
Storage Conditions 4°C
Shipping Conditions Gel Pack
USAGE For Research Use Only! Not For Use in Humans.

Details

Acute phase-regulated receptor involved in clearance and endocytosis of hemoglobin/haptoglobin complexes by macrophages and may thereby protect tissues from free hemoglobin-mediated oxidative damage. May play a role in the uptake and recycling of iron, via endocytosis of hemoglobin/haptoglobin and subsequent breakdown of heme. Binds hemoglobin/haptoglobin complexes in a calcium-dependent and pH-dependent manner. Exhibits a higher affinity for complexes of hemoglobin and multimeric haptoglobin of HP*1F phenotype than for complexes of hemoglobin and dimeric haptoglobin of HP*1S phenotype. Induces a cascade of intracellular signals that involves tyrosine kinase-dependent calcium mobilization, inositol triphosphate production and secretion of IL6 and CSF1. BioVision’s QuickDetect™ CD163 ELISA kit is a sandwich ELISA assay for measuring the quantity of human CD163 in serum, plasma and cell culture and other biological fluid. The entire process can be performed within 90 minutes with only a single incubation and a wash step respectively, resulting in fewer handling steps, fewer errors and more consistent results.


Why buy BioVision Products?
Innovation
Affordability
Global Presence
Technical Support
BioVision aims to provide our customers innovative tools for accelerating drug discovery and biological research. BioVision offers >8,000 products including the most comprehensive array of assay kits for key targets in Metabolic pathways.
BioVision is committed to providing the highest quality products at a competitive price.
We have a broad network of global distributors who are ready to address your research needs and ensure fast delivery.
Our highly trained Technical Support team provides comprehensive product support and is dedicated to resolving your issues quickly and efficiently.
What is a sandwich ELISA?
The analyte of interest in a sample is "sandwiched" by an immobilized capture antibody coated on the plate and a labeled antibody used for detection. Sandwich ELISA is very sensitive. The density of color is proportional to the amount of analyte captured from the samples and can be quantified when compared with standard curve.
What is a competitive ELISA?
The endogenous unlabled analyte of interest in a sample is "competed" by the exogenous labeled antigen coated on the plate for a limited amount of antibody binding sites. Therefore, the lower signal indicate higher concentration of the analyte.
What dilution range should I use for the samples?
A preliminary experiment is always recommended when working with new samples and ELISA Kits to ensure all results fall within the detection range.
Can I extend the standard curve (in either direction)?
Extended standard curve is not recommended by BioVision.
What type of software is needed to graph a 4-parameter or 5-parameter curve?
SoftMax Pro by Molecular Devices, SigmaPlot® by Systat Software Inc., or others can be used for this purpose.
Low absorbance; No signal
Target present below detection limits of assay
Reduce dilution factor to increase sample concentration  (Preliminary experiment is recommended for optimal dilution range)
Incompatible sample types
Use samples recommended by the kit
Standard dilutions are unstable
Use fresh standard dilutions
Buffers/substrates are contaminated
Use new buffers and substrates
Insufficient incubation time 
Extend incubation time
Cover or seal plates during all incubation 
Plate washings too vigorous
Pipette wash buffer gently or make sure the automatic wash system has correct pressure
Wells scratched with pipette or washing tips
Use caution when dispensing and aspirating into and out of wells
Incubation temperature too low
Ensure incubations are carried out at temperatures recommended by the manual
Reagents are cold
Bring all reagents to room temperature before each assay
Plate read at incorrect wavelength

Ensure the plate reader is set at the correct wavelength recommended on the protocol

High Background
Insufficient washing
Increase number of washes; Add a 30 second soaking step in between washes;  at the end of each washing step, invert plate on absorbent tissue to completely drain all wells
Residual buffer/substrates remain in wells
Remove all residual buffer and substrates at the end of each wash
Too much HRP-Streptavidin
Centrifuge and mix the substrate vial before use to avoid percipitation and inconsistent HRP concentration
Excessive incubation time
Reduce incubation time
Substrate incubation carry out in light or wait too long to read plate after adding stop solution
Avoid light during incubation and read signals immediately after adding stop solution
Standards improperly reconstituted or diluted
Briefly spin vials before opening; inspect for undissolved material after reconstitution
Poor standard curve
Standard improperly diluted
Confirm dilutions are made correctly
Standard dilutions are unstable
Use freshly diluted standards
Pipetting error 
use calibrated pipette and stay consistent between each pipette
Curve doesn't fit scale
Try plotting using different scales  (log-log, 4 parameter logistic, 5 parameter etc.)
Large intra-coefficient of variation (CV)
Bubbles in wells
Avoid bubbles during experiment, eliminate all bubbles prior to reading
Insufficient washing
Increase number of washes; add a 30 second soak step in between washes;  at the end of each washing step, invert plate on absorbent tissue to completely drain all wells
Wells not washed equally/thoroughly
Use calibrated (automated) multi-channel pipette for consistent handling, make sure all parts of the automated washer are unobstructed
Edge effects
Seal the plate completely during incubation
Incomplete reagent mixing
Make sure all reagents are mixed thoroughly during each step
Inconsistent sample preparation or storage
Consistent sample preparation. Optimize sample storage conditions and minimize freeze and thaw cycles
Stacked plates
Avoid stacking plates during incubation
Plate sealers not used or reused
Cover assay plate with plate sealer, use a fresh sealer each time to prevent contamination
Large inter-coefficient of variation (CV))
Insufficient washing
Increase number of washes; add a 30 second soak step in between washes;  at the end of each washing step, invert plate on absorbent tissue to completely drain all wells
Inconsistent incubation temperature
Make sure to follow recommended incubation temperatures. Avoid fluctuations in temperature due to environmental conditions.
Plate sealers not used or reused
Cover assay plate with plate sealer, use a fresh sealer each time to prevent contamination